Abbott Diabetes Care Announces Recall of Certain Lots of Blood Glucose Test Strips in the United States and Puerto Rico

Dec 22, 2010

This press release is an announcement submitted by Abbott Diabetes Care, and was not written by Diabetes Health.

Abbott Diabetes Care today announced that it has initiated a recall of 359 lots (approximately 359 million strips) of Precision Xtra®, Precision Xceed Pro®, MediSense® Optium, Optium, OptiumEZ and ReliOn® Ultima Blood Glucose Test Strips in the United States and Puerto Rico.

The test strips are used with Abbott's Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium and OptiumEZ blood glucose monitoring systems. ReliOn Ultima test strips are used with the ReliOn Ultima blood glucose monitoring system. The blood glucose monitors are not being recalled and customers can continue to use the blood glucose monitors.

The affected test strips may give falsely low blood glucose results, which can lead users to try to raise their blood glucose when it is unnecessary and to fail to treat elevated blood glucose due to a falsely low reading. The problem appears to be related to longer than expected blood fill times. This may be dependent on the age of the strips and if the strips have been stored in (or exposed to) higher temperatures (above 72F and not to exceed 86F) for an extended period of time.

Customers can check if they have test strips from the affected lots by visiting www.precisionoptiuminfo.com and looking up their product lot number.

Customers who have test strips from these affected lots should discontinue use of the product. Abbott Diabetes Care will replace affected test strips at no charge. Abbott Diabetes Care initiated this recall following a routine internal quality review that indicated certain lots of Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium, OptiumEZ and ReliOn Ultima blood glucose test strips may experience longer than normal blood fill times which may cause falsely low blood glucose readings.

Customers are advised to contact their health care provider to determine testing options if they do not have any unaffected test strips to use.

However, if consumers must use affected test strips while in the process of obtaining new strips, they should:

  • Check the amount of time it takes for their blood glucose meter to start the "countdown" after they first apply blood to the test strip. They should start timing immediately after blood first makes contact with the test strip. If the meter takes longer than five seconds to start the countdown, that test strip is affected and the result should not be used. Users should check the time for each test strip they use because all of the strips in a package may not be affected to the same degree. If any reading appears lower than they would expect or does not seem to correlate with the way they are feeling, the user should contact their health care provider immediately.
  • If they do not immediately have access to unaffected strips and are unable to test their blood sugar, users need to be aware of symptoms of high blood sugar (hyperglycemia) and low blood sugar (hypoglycemia).
    • Symptoms of high blood sugar include excessive thirst, excessive urination, blurred vision, weakness, nausea, vomiting, and abdominal pain. If a user is experiencing any of these symptoms or is not feeling well, they should contact their health care professional immediately.
    • Symptoms of low blood sugar may include trembling, excessive sweating, weakness, hunger, confusion, and headache. Some individuals may have no symptoms at all before they develop unconsciousness or seizures. It is important to treat low blood sugars promptly to avoid loss of consciousness or a seizure. If a user is unable to obtain unaffected strips, they should contact their health care provider for advice on how to treat these symptoms before they occur.

The recommendation is that hospitals and user facilities stop using the recalled test strips and follow the steps below:

1.     Use test strips from unaffected lots.

2.     If the health care facility does not have any test strips from unaffected lots but has an alternative point-of-care blood glucose testing system, begin using the alternative system until new, unaffected test strips can be obtained.

If the health care facility does not have any test strips from unaffected lots or immediate access to an alternative point-of-care blood glucose testing system, health care providers should take the following steps:

  • Verify any critical glucose test results (e.g., test results that may be used to adjust insulin therapy in vulnerable patient populations) generated on the Precision Xceed Pro Blood Glucose Monitoring System using a central laboratory blood glucose method. Clinical judgment should be applied when deciding whether to act on results prior to verification.
  • Verify any Precision Xceed Pro Blood Glucose Monitoring System results that do not match a patient's symptoms, or seem unexpected for the patient's clinical status, using a central laboratory blood glucose method.
  • Monitor the amount of time it takes for the Precision Xceed Blood Glucose Meter to start the "countdown" after blood is first applied to the test strip. If the amount of time exceeds five seconds, discard that test strip immediately to prevent the risk of receiving a falsely low blood glucose result (if this occurs, health care providers should note the lot number of that test strip, discard all the test strips from that lot, and notify Abbott Diabetes Care immediately).

Customers can check if they have test strips from the affected lots by visiting www.precisionoptiuminfo.com and looking up their product lot number.

Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium, OptiumEZ and ReliOn Ultima customers can also get information by calling Abbott Diabetes Care customer service at 1-800-448-5234 (English) and 1-800-709-7010 (Espanol).

The website and customer service phone numbers listed above will also provide customers with information about returning affected test strips.

Unaffected Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium, OptiumEZ and ReliOn Ultima Blood Glucose Test Strip lots may be used by customers to test their blood glucose as directed by their health care providers.

Abbott Diabetes Care has reported this situation to the U.S. Food and Drug Administration and is working to make the necessary corrective actions to prevent this situation from recurring.

***

Source:

Abbott Diabetes Care


Categories: Blood Glucose, Blood Sugar, Diabetes, Diabetes, Food, Insulin, Low Blood Sugar, Products



You May Also Be Interested In...


Comments

Posted by cde on 30 December 2010

After I sent the list of lot numbers of the affected strips to all my clients, at least one has called because he does indeed have strips from an affected lot of Medisense Optium XCEED strips.

Therefore, I believe that some of these strips have been sold in Mexico, also, and not just in US and Puerto Rico.

The address of the list only is:
http://www.precisionoptiuminfo.com/img/Lot-Numbers.pdf

Dr. Stan De Loach
México, Distrito Federal
http://www.diabetes-safari.com


Add your comments about this article below. You can add comments as a registered user or anonymously. If you choose to post anonymously your comments will be sent to our moderator for approval before they appear on this page. If you choose to post as a registered user your comments will appear instantly.

When voicing your views via the comment feature, please respect the Diabetes Health community by refraining from comments that could be considered offensive to other people. Diabetes Health reserves the right to remove comments when necessary to maintain the cordial voice of the diabetes community.

For your privacy and protection, we ask that you do not include personal details such as address or telephone number in any comments posted.

Don't have your Diabetes Health Username? Register now and add your comments to all our content.

Have Your Say...


Username: Password:
Comment:
©1991-2013 Diabetes Health | Home | Privacy | Press | Advertising | Help | Contact Us | Donate | Sitemap

Diabetes Health Medical Disclaimer

The information on this site is not intended or implied to be a substitute for professional medical advice, diagnosis or treatment. All content, including text, graphics, images, and information, contained on or available through this website is for general information purposes only. Opinions expressed here are the opinions of writers, contributors, and commentators, and are not necessarily those of Diabetes Health. Never disregard professional medical advice or delay seeking medical treatment because of something you have read on or accessed through this website.